Conference Publication Details
Mandatory Fields
Temko, A;Korotchikova, I;Manaane, W;Lightbody, G;Boylan, G
BIOSIGNALS 2010: PROCEEDINGS OF THE THIRD INTERNATIONAL CONFERENCE ON BIO-INSPIRED SYSTEMS AND SIGNAL PROCESSING
VALIDATION OF AN AUTOMATED SEIZURE DETECTION SYSTEM ON HEALTHY BABIES Histogram-based Energy Normalization for Montage Mismatch Compensation
2010
January
Validated
1
()
Optional Fields
312
317
Seizures in newborn babies are commonly caused by problems such as lack of oxygen, haemorrhage, meningitis, infection and strokes. The aim of an automated neonatal seizure detection system is to assist clinical staff in a neonatal intensive care unit to interpret the EEG. In this work, the automated neonatal seizure detection system is validated on a set of healthy patients and its performance is compared to the performance obtained on sick patients reported previously. The histogram-based energy normalization technique is designed and applied to EEG signals from healthy patients to cope with montage mismatch. The results on healthy babies compares favourably to those obtained on sick babies. Several useful observations are made which were not possible to obtain by testing on sick babies only such as a practically useful range of probabilistic thresholds, minimum detection duration restriction, and an influence of the database statistics on the system performance.
Grant Details